Products

IECEx Approved RTD Pt100 Sensors

with Stainless Steel Head and Fixed Process Thread

Our IECEx approved flexible mineral insulated RTD Pt100 utilise, as standard, detector elements with a resistance of 100 ohms @ 0°C and a fundamental interval of 38.5 ohms to IEC 60751. Single and duplex elements are available.
  • Available in a wide range of protection concepts for Zones 0, 1 & 2 use. See table below for details
  • Temperature classification T6-T1, see here for stand-off requirements
  • Available in B, A, 1/3, 1/5 and 1/10 grade accuracies
  • Sheath diameters from 0.188" to 0.313"
  • 316 seamless stainless steel sheath (other grades/materials available)
  • Wide operating temperature range of -148°F to +1112°F
  • UKAS calibration available.
Available with the following approvals:

Protection Type Description Zone
Ex nA IIC Gc Non-Sparking (Gas) 2
Ex d IIC Gb Flameproof (Gas) 1
Ex ia IIC Ga / IIIC Da Intrinsically Safe (Gas/Dust) 0/20
Ex tb IIIC Db Enclosure (Dust) 21
Ex tc IIIC Dc Enclosure (Dust) 22



Enquiry Form

1. No. of Elements


2. Protection Concept


Concept Type Select
Ex d Flameproof/Explosion Proof
Ex ia Intrinsically Safe
Ex nA Non-Sparking
Ex tb Enclosure
Ex tc Enclosure

Your chosen concept is:


Ex d Flameproof/Explosion Proof – Category 2, Zones 1,2 (Gas)
• Ex d IIC T6...T1 Gb Ta = –40°C to 60°C
• Standard EN IEC 60079-1 refers
• Designed to contain an explosion and quench the flame
• Sensors incorporate a suitable heavy duty enclosure or threaded fitting with tight seals

Your chosen concept is:


Ex e Increased Safety – Category 2, Zones 1,2 (Gas)
• Ex e IIC T6...T1 Gb Ta = –40°C to 60°C
• Standard EN IEC 60079-7 refers
• Design ensures no arcs, sparks or hot surfaces that could ignite an explosive mixture with a minimum IP54 enclosure rating
• Sensors incorporate a suitable heavy duty enclosure or threaded fitting with tight seals

Your chosen concept is:


Ex ia Intrinsically Safe – Category 1, Zones 0,1,2 (Gas) & 20,21,22 (Dust)
• Ex ia IIC T6...T1 Ga Ta = –40°C to 60°C
• Ex ia IIIC T80°C…T440°C Db
• Standard EN IEC 60079-11 refers
• Designed to limit the energy of arcs or sparks that could potentially ignite an explosive mixture
• Sensors used in this area must be wired through a suitably approved barrier
• Optional approved 4 to 20mA transmitters available

Your chosen concept is:


Ex nA Non-Sparking – Category 3, Zone 2 (Gas)
• Ex nA IIC Gc T6…T2
• Standard EN IEC 60079-15 refers
• No arcs, sparks or hot surfaces

Your chosen concept is:


Ex tb Enclosure – Category 2, Zones 21, 22 (Dust)
• Ex tb IIIC T80°C…T440°C Db
• Standard EN IEC 60079-31 refers
• Protects against combustible and conductive dusts, rugged tight enclosures are used

Your chosen concept is:


Ex tc Non-Sparking – Category 3, Zone 22 (Dust)
• Ex tc IIIC T85°C...T450°C Dc
• Standard EN IEC 61241 refers
• Protects against combustible dusts
• Sensors incorporate a suitable heavy duty enclosure with IP6X seal

3. Number of Wires


4. Sheath Diameter


5. Sheath Length


 inches*

6. Tolerance of Element


Grade

Accuracy

@ 0ºC

Accuracy

@ 100ºC

Select
B +/- 0.30ºC +/- 0.80ºC
A +/- 0.15ºC +/- 0.35ºC
1/3 +/- 0.08ºC +/- 0.19ºC
1/5 +/- 0.05ºC +/- 0.15ºC
1/10 +/- 0.03ºC +/- 0.12ºC

7. Process Connection Thread


8. Optional 4 to 20mA Transmitter


9. Quantity Required


Contact Details


Company Name


Your Name


Email


Telephone


Additional Information


If your enquiry is for more than one style of sensor, please email us with your requirements or use the box below



Send Enquiry  
BACK TO TOP
IECEx Thermocouples with Pot Seal

IECEx Thermocouples with Die Cast Alloy Head

IECEx Thermocouples with Die Cast Alloy Head and Process Connection

IECEx Thermocouples with St. / Steel Head

IECEx Thermocouples with St. / Steel Head and Process Connection

IECEx Thermocouples with Dual Entry Head

IECEx Thermocouples with Dual Entry Head and Process Connection

IECEx Thermocouples with Spring loaded terminal block

IECEx Thermocouples with Terminal Entry Gland

IECEx Thermocouples with Terminal Entry Gland and Process Connection

IECEx Spring Loaded Thermocouples

IECEx Thermocouple Feedthroughs
IECEx RTD Pt100 Sensors with Pot Seal

IECEx RTD Pt100 Sensors with Die Cast Alloy Head

IECEx RTD Pt100 Sensors with Die Cast Alloy Head and Process Connection

IECEx RTD Pt100 Sensors with St. / Steel Head

IECEx RTD Pt100 Sensors with St. / Steel Head and Process Connection

IECEx RTD Pt100 Sensors with Dual Entry Head

IECEx RTD Pt100 Sensors with Dual Entry Head and Process Connection

IECEx RTD Pt100 Sensors with Spring loaded terminal block

IECEx RTD Pt100 Sensors with Terminal Entry Gland

IECEx RTD Pt100 Sensors with Terminal Entry Gland and Process Connection

IECEx Spring Loaded RTD Pt100 Sensors
IECEx 4 to 20mA Indicators

IECEx Temperature Transmitters

IECEx Compression Fittings

IECEx Terminal Heads

IECEx Pressure and Vacuum Feedthroughs



© 2006- TC Measurement and Control, Inc, PO Box 65, Hillside, IL 60162
Tel.: 877 249 1777, Fax: 708 449 0777, Email: info@tc-inc.com







Wiring Diagram Sheath Materials